Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.
Microscan Microstrep Panel (March 08, 2013)
- Starting date:
- March 08, 2013
- Posting date:
- March 08, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Product Safety
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-26451
Reason
There is an issue of false skipped wells and false susceptible misreads with s. agalactiae affecting multiple antimicrobial agents on Microscan MSP1, MSP5, MF5J and MF7J panels processed in a WalkAway system. The effect is the potential for false susceptible results for s. agalactiae if the walkaway system reads are not verified. This issue is not panel lot specific nor is it related to a particular model of WalkAway instrumentation. No other streptococcal species are affected.
Affected products
Microscan Microstrep Panel (2013-03-08)
Lot or serial number
All lots
Model or catalog number
10444757, B1027-201
Companies
- Manufacturer
- Not Applicable