CT/MR Fletcher Set (2014-03-17)
- Starting date:
- March 17, 2014
- Posting date:
- April 14, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-39005
Recalled Products
CT/MR Fletcher Set
Reason
The CT/MR Fletcher set is designed for gynaecological brachytherapy (Radiation therapy) and mainly intented for the treatment of the cervix. During tests done at a customer site, it has been revealed by the customer's physicist that there is a slight deviation in the accuracy of the dwell position of the source.
Affected products
CT/MR Fletcher Set
Lot or serial number
Not applicable
Model or catalog number
SET0109
Companies
- Manufacturer
-
Eckert & Ziegler Bebig GMBH