This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Vitros Chemisty Products HSCRP Reagent (2014-05-12)

Starting date:
May 12, 2014
Posting date:
June 5, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39875

Recalled Products

Vitros Chemisty Products HSCRP  Reagent 

Reason

Calibrator values were released using a reference method that was undergoing validation for a new lot of European Reference Material (ERM). This validation process was not complete prior to calibrator value assignment for Vitros chemistry products HSCRP reagent (Product Code 6801739), Generation (Gen) 28.

Affected products

Vitros Chemisty Products HSCRP  Reagent (

Lot or serial number

1508-28-3487
 

Model or catalog number

6801739
 

Companies
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES