ITS4850A Intellivue Telemetry System - Moinitor (2014-05-20)
- Starting date:
- May 20, 2014
- Posting date:
- June 12, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-39997
Recalled products
- ITS4850A Intellivue Telemetry System - Moinitor
Reason
Philips has become aware that IntelliVue MX40 System (part numbers 865350, 865351, 865352) with software revisions B.00 or B.01 may not resume alarming when the alarms are paused and the pause timer ends. This issue will only occur when Teh MX40 device is in monitor mode connected to a Philips IntelliVue Information Center iX (PIIC iX) that is not in a specific default configuration.
Affected products
A. ITS4850A Intellivue Telemetry System - Moinitor
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- MX40
Companies
- Manufacturer
-
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES