Axiom Artis DTA,Artis Zeego and Artis Zee- Multi-purpose, Floor, Celing and Biplane (2014-06-11)
- Starting date:
- June 11, 2014
- Posting date:
- June 24, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-40179
Recalled products
- Axiom Artis DTA
- Artis Zeego
- Artis Zee Multi-purpose
- Artis Zee Floor
- Artis Zee Ceiling
- Artis Zee Biplane
Reason
The hand switch to release a x-ray image at the patient table is equipped with a cable extension inside the table. A problem may occur if the connection between cable connection and hand switch is contaminated with an extraordinary large amount of fluid. In a worst case situation, uncontrolled release of a x-ray image may happen. If this situation occurs, the system terminates uncontrolled x-ray after a short period.
Affected products
A. Axiom Artis DTA
Lot or serial number
- 55402
- 55433
Model or catalog number
- 7008605
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
B. Artis Zeego
Lot or serial number
- 160109
- 160354
- 160423
Model or catalog number
- 10280959
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
C. Artis Zee Multi-purpose
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- 10094139
- 7555365
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
D. Artis Zee Floor
Lot or serial number
- 135367
- 136099
- 136104
- 136226
Model or catalog number
- 10094135
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
E. Artis Zee Ceiling
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- 10094137
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
F. Artis Zee Biplane
Lot or serial number
- 153146
- 153223
- 153275
- 153369
- 153372
- 153810
- 153854
- 153859
- 44274
Model or catalog number
- 10094141
- 7728392
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY