This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Qube Compact Monitor (2014-06-06)
- Starting date:
- June 6, 2014
- Posting date:
- June 24, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-40183
Recalled products
- Qube Compact Monitor
Reason
Potential issue with Qube compact monitors that are "docking station ready" - that have had a port cover plate removed allowing the monitor to be used with the optional Qube docking station. If a docking station ready Qube monitor is powered up undocked, connected to the network via an ethernet hardwire connection, and then disconnected from the ethernet hardwire connection and inserted into a Qube docking station, a switch condition will cause the monitor to lose the network connection. All bedside alarms will continue to function normally. No one has been injured as a result of this issue.
Affected products
A. Qube Compact Monitor
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- 91390
Companies
- Manufacturer
-
Spacelabs Healthcare Inc.
35301 SE Center Street
Snoqualmie
98065
Washington
UNITED STATES