Artis Zeego and Artis Zee Multi-purpose, Floor, Celing and Biplane (2014-06-12)
- Starting date:
- June 12, 2014
- Posting date:
- June 24, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-40185
Recalled products
- Artis Zeego
- Artis Zee Floor
- Artis Zee Ceiling
- Artis Zee Biplane
Reason
Siemens is informing all users of Artis Zee and Artis Zeego systems that under certain circumstances, the release of radiation can become blocked unnecessarily. This situation comes about after the system has booted up, and after an error with the large display occurs that causes the system to block the release of radiation. After this, the system remains in a state where radiation cannot be released, even if the error that triggered this state was only a temporary one, and the error condition itself goes away after a few seconds (such as a brief communication error).
Affected products
A. Artis Zeego
Lot or serial number
- 160810
- 160846
Model or catalog number
- 10280959
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
B. Artis Zee Floor
Lot or serial number
- 136972
- 137167
Model or catalog number
- 10094135
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
C. Artis Zee Ceiling
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- 10094137
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen, Germany
80333
D. Artis Zee Biplane
Lot or serial number
- 153810
- 153854
- 153859
- 154569
- 154679
- 154691
- 154693
Model or catalog number
- 10094141
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY