Insight and Ultra-Speed (2014-07-02)
- Starting date:
- July 2, 2014
- Posting date:
- July 10, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-40407
Recalled Products
A) Insight
B) Ultra-Speed
Reason
Carestream has identified an incorrect orientation of the indicator dot on dental intraoral film packet labeling of some lots of insight, item 811 0785 (size 0 packet), ib-01, and some lots of ultra-speed, item 834 8658 (size 0 packet), df-48. The indicator dot on the film packet's outer printed paper labeling was incorrectly positioned relative to the location of the raised dot on the film. This issue may cause some confusion for the dental practitioner if the processed film is placed on the view box by referencing the film's raised dot. The result is the image orientation will be rotated 180 degrees. The remaining content on the label is correct. Image quality is not affected by this issue.
Affected products
A) Insight
Lot or serial number
538-013-03
539-012-01
545-012-01
546-012-01
Model or catalog number
811 0785
Companies
- Manufacturer
-
Carestream Health Inc.
150 Verona Street
Rochester
14608
New York
UNITED STATES
B) Ultra-Speed
Lot or serial number
333-012-01
333-012-02
333-012-03
343-015-01
Model or catalog number
834 8658
Companies
- Manufacturer
-
Carestream Health Inc.
150 Verona Street
Rochester
14608
New York
UNITED STATES