Dialog+ SW 9.XX Two Pump Machines (incl. HDF online) (2014-11-17)
- Starting date:
- November 17, 2014
- Posting date:
- November 26, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-42603
Recalled Products
Dialog+ SW 9.XX Two Pump Machines (incl. HDF online)
Reason
There is a risk of air infusion into the arterial line if the valve block level regulation and/or diaphragm pump is installed incorrectly during technical service. The correct functionality of manual regulation must be tested after service. The notification affects only units in which the valve block has been exchanged.
Affected products
A) Dialog+ SW 9.XX Two Pump Machines (incl. HDF online)
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
710201CV
Companies
- Manufacturer
-
B. Braun Avitum AG
Schwarzenberger Weg 73-79
Melsungen
34212
GERMANY