Health product recall

Dialog+ SW 9.XX Two Pump Machines (incl. HDF online) (2014-11-17)

Starting date:
November 17, 2014
Posting date:
November 26, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-42603

Recalled Products

Dialog+ SW 9.XX Two Pump Machines (incl. HDF online)

Reason

There is a risk of air infusion into the arterial line if the valve block level regulation and/or diaphragm pump is installed incorrectly during technical service. The correct functionality of manual regulation must be tested after service. The notification affects only units in which the valve block has been exchanged.

Affected products

A) Dialog+ SW 9.XX Two Pump Machines (incl. HDF online)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

710201CV

Companies
Manufacturer
B. Braun Avitum AG
Schwarzenberger Weg 73-79
Melsungen
34212
GERMANY