This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Archivé - Important Safety Information on the Interactions between Active Implantable Medical Devices and Systems and other Medical Devices - Update
- Starting date:
- December 19, 2005
- Posting date:
- December 20, 2005
- Type of communication:
- Notice to Hospitals
- Subcategory:
- Medical Device
- Source of recall:
- Health Canada
- Audience:
- Healthcare Professionals
- Identification number:
- RA-170001080
Notice about Health Canada advisories
Notice to Hospitals - Health Canada Issued Important Safety Information on Active Implantable Medical Devices and Systems
December 19, 2005
To: Hospital Chief of Medical Staff
Please distribute to the relevant Departments of Physiotherapy, Occupational Therapy, Radiology, Rehabilitative Medicine, Surgery, Neurosurgery, Orthopedic Surgery, Biomedical Engineering, Emergency Medicine, Pediatrics, Geriatrics, Internal Medicine, Nursing, Dentistry, Intensive Care and other involved professional staff and post this NOTICE in your institution.
Subject: Interactions between active implantable medical devices and systems and other medical devices
This advisory updates previous Notices to Hospitals (October 5, 2005 and October 31, 2005) on this topic, based on recent information provided to Health Canada by device manufacturers and health care providers. The change is highlighted in the table.
Patients with active implantable medical devices and systems (AIMDS) are at risk of injury from interactions between the AIMDS and other medical devices used for diagnostic or therapeutic purposes. Such interactions could damage some AIMDS resulting in loss of therapy from these units, and could cause tissue damage resulting in severe injury or death.
On April 2, 2003, Heath Canada issued a Notice to Hospitals entitled "Important safety information regarding DIATHERMY THERAPY in patients with implanted leads and implanted systems with leads".Footnote 1 It warned about the risks of tissue overheating, serious injury and death in this group of patients when subject to shortwave (radiofrequency) or microwave diathermy therapy. Since then, Health Canada has become aware of further international reports of serious injury resulting from interactions between AIMDS and other medical devices.Footnote 2
In 2003, the French Health Products Safety Agency, Agence française de sécurité sanitaire des produits de santé (AFSSAPS), assembled a working group to analyse the recommendations of AIMDS manufacturers and make recommendations to health care professionals who perform certain diagnostic and therapeutic procedures on patients with AIMDS. The recommendations of the group were released by AFSSAPS in [NJS_FILE:3311bdad-28b4-40a2-8077-a66dbc3cbc0f:5c647379-41d6-40fd-91d4-571527c2ed9a].Footnote 3
Procedure | Type of AIMDS | ||
---|---|---|---|
Implanted Cardiac Pacemaker | Implanted Cardiac Defibrillator | Implanted Neurostimulator | |
* This type only of MRI is allowed if both of the following conditions are met:
|
|||
Shortwave and Microwave Diathermy | Absolute contraindication | Absolute contraindication | Absolute contraindication |
Therapeutic Ultrasound Diathermy | Relative contraindication | Relative contraindication | Absolute contraindication |
Electromagnetic Stimulation | Relative contraindication | Relative contraindication | Absolute contraindication |
MRI with transmit/receive radio-frequency (RF) head coilFootnote 4 | Absolute contraindication | Absolute contraindication | Relative contraindication* |
MRI (all other) | Absolute contraindication | Absolute contraindication | Absolute contraindication |
Electrosurgery | Relative contraindication | Relative contraindication | Relative contraindication |
External Defibrillation | Relative contraindication | Relative contraindication | Relative contraindication |
Radiotherapy | Relative contraindication | Relative contraindication | Relative contraindication |
Lithotripsy or Ultrasound Therapy | Relative contraindication | Relative contraindication | Relative contraindication |
Fluoroscopy or other X-ray procedures | No contraindication | No contraindication | No contraindication |
Echography | No contraindication | No contraindication | No contraindication |
Definitions of Levels of Risk Management - AIMDS
No contraindication: The risk of the procedure remains unchanged due to the presence of an AIMDS.
Relative Contraindication: In some circumstances the benefits of the procedures may outweigh its risks. A risk-benefit analysis of the proposed medical procedure, including careful consideration of the alternatives, should be undertaken. Consultation with the referring physician and implanting/monitoring physician should be considered and the labelling of relevant devices (including warnings/precautions) should be reviewed. The exact risks to patients and precautions to be taken in each circumstance are outlined in detail in the AFSSAPS document.Footnote 3 In general, if a decision is taken to go ahead with a medical procedure that requires precautions, Health Canada recommends the following:
- if possible, deactivate the AIMDS
- use the lowest device strength compatible with an acceptable diagnostic or therapeutic outcome and keep the path of device output as far from the AIMDS as possible
- monitor the patient closely during the medical procedure
- stop the procedure immediately in the case of an adverse incident
- verify the continued proper function of the AIMDS during and after the medical procedure.
Absolute Contraindication: The procedure should not be used because the risk always outweighs the benefit.
The risk of interactions between AIMDS and other medical devices used for diagnostic or therapeutic purposes depends on many parameters. Due to the diversity and complexity of the devices in question, Health Canada wishes to pass along these general guidelines but cannot give all-encompassing recommendations that cover all possible clinical scenarios. The recommendations outlined in this document are based on the information that has been received so far, and Health Canada will keep monitoring this evolving safety issue. The recommendations will be updated as more information will become available.
Health Canada depends on health care professionals to report adverse incidents related to medical devices. Any serious and/or unexpected adverse incident related to medical devices should be reported to Health Canada at the following address:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Tel: The Inspectorate Hotline 1-800-267-9675
The Reporting Form and Guidelines can be obtained from the Health Canada web site or in The Canadian Compendium of Pharmaceuticals and Specialties.
For other inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate (MHPD)
MHPD_DPSC@hc-sc.gc.ca
Tel/Tél: -613-954-6522
Fax/Télé: 613-952-7738
References
- Footnote 1
-
Important safety information regarding DIATHERMY THERAPY in patients with implanted leads and implanted systems with leads. Notice To Hospitals. Health Products and Food Branch, Health Canada. April 2, 2003.
- Footnote 2
-
FDA Public Health Notification: MRI-Caused Injuries in Patients with Implanted Neurological Stimulators. Center for Devices and Radiological Health, Food and Drug Administration. May 10, 2005.
- Footnote 3
-
Interactions entre dispositifs médicaux implantables actifs et dispositifs médicaux. [NJS_FILE:3311bdad-28b4-40a2-8077-a66dbc3cbc0f:f2ed8c59-1053-4c1f-8beb-d7d42f9df4f9]. February 2005.
- Footnote 4
-
MRI and Activa Therapy-To the physician, Medtronic Inc. (2005)