Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.
Ecc. O. Integrated Mini Bypass System with Preconnected Tubing, Primo2x Oxygenator w Integrated Hardshell Venous Reservoir, Synthesis Oxygenator w Heat Exchanger Arterial Filter and Venous Reservoir, D903 Avant 2 Adult Hollow Fiber Oxygenator, Synergy Adult Integrated Mini Bypass System, D905 EOS Pediatric Hollow Fiber Oxygenator, D905 Module Pediatric Hollow Fiber Oxygenator, Apex HP Phisio Smart Hollow Fiber Membrane Oxygenator , I) D903 Oxygenator Perfusion Tubing Pack Phisio or Non Phisio Coated w Oxygenator, Primo2X Oxygenator Perfusion Tubing Pack Phisio or Non Phisio Coated w Oxygenator, Synthesis Oxygenator Perfusion Tubing Pack Phisio or Non Phisio Coated w Oxygenator, Synthesis Oxygenator Perfusion Tubing Pack Phisio Coated w Oxygenator, D901 Oxygenator Perfusion Tubing Pack Phisio Coated w Oxygenator, D902 Oxygenator Perfusion Tubing Pack Phisio Coated w Oxygenator and D905 EOS Oxygenator Perfusion Tubing Pack Phisio Coated w Oxygenator (December 08, 2011)
- Starting date:
- December 08, 2011
- Posting date:
- January 09, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Product Safety
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69647
Reason
Sorin Group Italia has received customer complaints reporting a leak between the blood and water compartments of the heat exchanger used in the referenced oxygenators. The occurrence rate has been extremely low (less than 0.002%) and to date there have been no reports of any associated patient injuries.
Affected products
Ecc. O. Integrated Mini Bypass System with Preconnected Tubing
Lot or serial number
All lots
Model or catalog number
085339
Companies
- Manufacturer
- Not Applicable
Primo2x Oxygenator w Integrated Hardshell Venous Reservoir
Lot or serial number
All lots
Model or catalog number
050533
Companies
- Manufacturer
- Not Applicable
Synthesis Oxygenator w Heat Exchanger Arterial Filter and Venous Reservoir
Lot or serial number
All lots
Model or catalog number
050238, 050239
Companies
- Manufacturer
- Not Applicable
D903 Avant 2 Adult Hollow Fiber Oxygenator
Lot or serial number
All lots
Model or catalog number
5338
Companies
- Manufacturer
- Not Applicable
Synergy Adult Integrated Mini Bypass System
Lot or serial number
All lots
Model or catalog number
074622000
Companies
- Manufacturer
- Not Applicable
D905 EOS Pediatric Hollow Fiber Oxygenator
Lot or serial number
All lots
Model or catalog number
050512
Companies
- Manufacturer
- Not Applicable
D905 Module Pediatric Hollow Fiber Oxygenator
Lot or serial number
All lots
Model or catalog number
050513
Companies
- Manufacturer
- Not Applicable
Apex HP Phisio Smart Hollow Fiber Membrane Oxygenator
Lot or serial number
All lots
Model or catalog number
050552
Companies
- Manufacturer
- Not Applicable
I) D903 Oxygenator Perfusion Tubing Pack Phisio or Non Phisio Coated w Oxygenator
Lot or serial number
I) All lots
Model or catalog number
I) 084423
Companies
- Manufacturer
- Not Applicable
Primo2X Oxygenator Perfusion Tubing Pack Phisio or Non Phisio Coated w Oxygenator
Lot or serial number
All lots
Model or catalog number
097794, 097891, AB0638
Companies
- Manufacturer
- Not Applicable
Synthesis Oxygenator Perfusion Tubing Pack Phisio or Non Phisio Coated w Oxygenator
Lot or serial number
All lots
Model or catalog number
097844, 098013, AB0350, AB0405
Companies
- Manufacturer
- Not Applicable
Synthesis Oxygenator Perfusion Tubing Pack Phisio Coated w Oxygenator
Lot or serial number
All lots
Model or catalog number
AB0097
Companies
- Manufacturer
- Not Applicable
D901 Oxygenator Perfusion Tubing Pack Phisio Coated w Oxygenator
Lot or serial number
All lots
Model or catalog number
082231
Companies
- Manufacturer
- Not Applicable
D902 Oxygenator Perfusion Tubing Pack Phisio Coated w Oxygenator
Lot or serial number
All lots
Model or catalog number
084590, 084592
Companies
- Manufacturer
- Not Applicable
D905 EOS Oxygenator Perfusion Tubing Pack Phisio Coated w Oxygenator
Lot or serial number
All lots
Model or catalog number
084424, 084591, 086499, 087582
Companies
- Manufacturer
- Not Applicable