Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.
Codman I.C.P. Express Monitoring System - Microsensor Kit (February 13, 2012)
- Starting date:
- February 13, 2012
- Posting date:
- April 02, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Product Safety
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72434
Reason
The hex wrench supplied with certain Codman ICP Skull Bolt Kits may not be made according to product specifications. This could result in a non-functional wrench that cannot be used to turn the set screw on the drill.
Affected products
Codman I.C.P. Express Monitoring System - Microsensor Kit(2012-02-13)
Lot or serial number
CMJB68, CMJC1J, CMLCJV, CMLBM1, CMNBZ4, CMLCJT, CMMBL6, CMKB2Z
Model or catalog number
82-6638