Gemstar Docking Station (2012-12-08)
- Starting date:
- December 8, 2012
- Posting date:
- March 18, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-38249
Recalled Products
Gemstar Docking Station
Reason
The Gemstar stations do not meet the Hospira product specification for output voltage. Under certain conditions, the potential exists for the pump to fail to power up while connected to the docking station. Moreover under a specific configuration, the pump will display error code 11/003 indicating excessive input voltage from the external sources. This code will stop infusion if the error occurs while an infusion is in progress.
Affected products
Gemstar Docking Station
Lot or serial number
All serial numbers
Model or catalog number
13075
13075-XX-YY
Companies
- Manufacturer
-
Hospira Inc.
275 North Field Drive
Lake Forest
60045
Illinois
UNITED STATES