Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.

Vitros 5.1 FS, 5600, and 4600 Systems (June 07, 2013)

Starting date:
June 07, 2013
Posting date:
June 07, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Product Safety
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-34985

Reason

Ortho-Clinical Diagnostics has announced a recall to inform customers of a change to the frequency of replacement for the CM/RT wear pads on Vitros 5600, 4600 and 5,1 FS systems.

CM/RT wear pads are a component in the CM/RT incubator. The CM/RT rotor revolves on three wear pads within the incubator. By design, there is a small gap between the rotor and the hot plate to allow for the transfer of heat. If the wear pads become too thin, it may cause the rotor to rub on the hot plate resulting in black dust within the incubator.

The impact of black dust may contribute to unplanned system down time. We have no data that suggests that black dust has any adverse effect on results.

Affected products

A. Vitros 5,1 FS Chemistry System

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

6801375, 6801890, 6802413, 6802445

Companies

Manufacturer
Not Applicable



B. Vitros 5,1 FS Chemistry System - Refurbished

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

6801375, 6801890, 6802413, 6802445

Companies

Manufacturer
Not Applicable



C. Vitros 5600 Integrated System

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

6801375, 6801890, 6802413, 6802445

Companies

Manufacturer
Not Applicable



D. Vitros 4600 Chemistry System

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

6801375, 6801890, 6802413, 6802445

Companies

Manufacturer
Not Applicable