Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.

SysMex Poch-100I System Analyzer (May 15, 2013)

Starting date:
May 15, 2013
Posting date:
May 15, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Product Safety
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-35191

Reason

In rare instances, the data printed on the thermal printer may omit a digit or decimal point in the results or sample identification number. This is isolated to the thermal printout, and does not affect the values displayed on the LCD screen, graphic printer or output to the host computer.

Affected products

SysMex Poch-100I System Analyzer (2013-05-13)

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

023-1861-8

Companies

Manufacturer
Not Applicable