Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.
SysMex Poch-100I System Analyzer (May 15, 2013)
- Starting date:
- May 15, 2013
- Posting date:
- May 15, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Product Safety
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-35191
Reason
In rare instances, the data printed on the thermal printer may omit a digit or decimal point in the results or sample identification number. This is isolated to the thermal printout, and does not affect the values displayed on the LCD screen, graphic printer or output to the host computer.
Affected products
SysMex Poch-100I System Analyzer (2013-05-13)
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
023-1861-8
Companies
- Manufacturer
- Not Applicable
- Date modified: