Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.

Endeavour Light System (2013-10-28)

Starting date:
October 28, 2013
Posting date:
November 29, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36929

Recalled Products

Endeavour Light System

Reason

If user fails to secure the cable to the temple arm of his/her loupes with the spiral temple clip provided with the device, strain on the cable (cable pull forces in excess of 7 pounds) where it interfaces to the headlight may result in an electrical open that prevents power from reaching the headlight LED.

Affected products

Endeavour Light System 

Lot or serial number

78100100 to 78113815

Model or catalog number

922003-X

Companies
Manufacturer
Kerr Corporation
190 - 3225 Derning Way
Middleton
53562
Wisconsin
UNITED STATES