Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.
The Ivent 201 Disposable Breathing Patient Circuit (2014-01-13)
- Starting date:
- January 13, 2014
- Posting date:
- February 11, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-37881
Recalled products
- The Ivent 201 Disposable Breathing Patient Circuit
Reason
Certain vital signs breathing circuits may be provided with 15 mm blue port caps as a loose component and/or 22 mm blue port caps attached to certain circuit ports. They are not intended to be part of the final circuit assembly. If the circuit assembly is used with the cap in place it will cause occlusion of gas flow.
Affected products
A. The Ivent 201 Disposable Breathing Patient Circuit
Lot or serial number
Not applicable
Model or catalog number
- M1171508
Companies
- Manufacturer
-
GE Medical Systems Israel Ltd.
Hasharon Industrial Park, Ornat Building
Kadima
0060920
ISRAEL