Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.

AU Chemistry Systems: ISE Cleaning Solution (2014-01-13)

Starting date:
January 13, 2014
Posting date:
February 11, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37893

Recalled Products

  1. AU Chemistry Systems: ISE Cleaning Solution

Reason

Recent customer feedback has indicated that in some cases, AUH1019 ISE cleaning solution may not be effectively cleaning AU ISE systems. This issue may be due to gradual decline in chlorine (Sodium hypochlorite) concentration over the shelf life of the product.

Affected products

A. AU Chemistry Systems: ISE Cleaning Solution

Lot or serial number

All lots  

Model or catalog number
  • AUH1019
Companies
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES