Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.

Dimension Vista System - Enzyme Calibrator (2014-01-20)

Starting date:
January 20, 2014
Posting date:
February 11, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37925

Recalled products

  1. Dimension Vista System - Enzyme Calibrator 

Reason

There have been reports of quality control (QC) shifts with pseudocholinesterase (PCHE) and imprecision during PCHE calibration with Dimension Vista enzyme 1 calibrator (ENZ 1 cal) lots 3FD034 and 3HD012. Investigation by Siemens confirms a loss of PCHE activity in the affected lots of enzyme 1 calibrator, causing an elevation of qc by as much as 80% and in some instances, imprecision of replicates during calibration.

Affected products

A. Dimension Vista System - Enzyme Calibrator 

Lot or serial number
  • 3FD034
  • 3HD012
Model or catalog number
  • KC310
Companies
Manufacturer
Siemens Healthcare Diagnostics Inc.