Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.
Immulite 2000 System - Osteocalcin Assay (2014-01-20)
- Starting date:
- January 20, 2014
- Posting date:
- February 11, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-37927
Recalled products
- Immulite 2000 System - Osteocalcin Assay
Reason
Siemens Healthcare Diagnostics has confirmed under recovery up to 50% in patient values with the Immulite 2000 / Immulite 2000 XPI Osteocalcin (l2KON2) kit lots 218, 219, 220 and 221. Quality control materials will not detect the bias.
Affected products
A. Immulite 2000 System - Osteocalcin Assay
Lot or serial number
- 218
- 219
- 220
- 221
Model or catalog number
- L2KON2
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM