Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.

Symbia T series Gamma Camera/CT Scanners (2014-08-06)

Starting date:
August 6, 2014
Posting date:
August 25, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41191

Recalled Products

Symbia T series Gamma Camera/CT Scanners 

Reason

Siemens has detected a rubber isolator with unfavourable aging properties. The function of this rubber isolator, which is situated within the rotate motor mounting assembly. It may degrade over time and may impact motor support.

Affected products

Symbia T series Gamma Camera/CT Scanners 

Lot or serial number

All lots.    

Model or catalog number
  • 10275007
  • 10275008
  • 10275009
  • 10275010
Companies
Manufacturer
Siemens Medical Solutions USA Inc.
Molecular Imaging Group,
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES