This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Steris 5085 and 5085SRT Surgical Table (2014-10-13)
- Starting date:
- October 13, 2014
- Posting date:
- October 22, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-41815
Recalled Products
A) Steris 5085 Surgical Table
B) Steris 5085SRT Surgical Table
Reason
Steris has learned through customer feedback and field service experience that the shroud cover for the hydraulic column may become misaligned and damaged as a result of extreme table articulation or excessive pressure on the side or base of the table. In most instances, the shroud damage is minor and does not affect the vertical up/down movement of the hydraulic column. In cases of severe shroud damage, the shroud sections may become misaligned such that vertical movement of the hydraulic column is impeded.
Affected products
A) Steris 5085 Surgical Table
Lot or serial number
0403809001 to 0425514019
Model or catalog number
ST01-410-X
Companies
- Manufacturer
-
Steris Corporation
2720 Gunter Park East
Montgomery
36109
Alabama
UNITED STATES
B) Steris 5085SRT Surgical Table
Lot or serial number
0403809001 to 0425514019
Model or catalog number
STO1-420-X
Companies
- Manufacturer
-
Steris Corporation
2720 Gunter Park East
Montgomery
36109
Alabama
UNITED STATES