Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.

Steris 5085 and 5085SRT Surgical Table (2014-10-13)

Starting date:
October 13, 2014
Posting date:
October 22, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41815

Recalled Products

A) Steris 5085 Surgical Table
B) Steris 5085SRT Surgical Table

Reason

Steris has learned through customer feedback and field service experience that the shroud cover for the hydraulic column may become misaligned and damaged as a result of extreme table articulation or excessive pressure on the side or base of the table. In most instances, the shroud damage is minor and does not affect the vertical up/down movement of the hydraulic column. In cases of severe shroud damage, the shroud sections may become misaligned such that vertical movement of the hydraulic column is impeded.

Affected products

A) Steris 5085 Surgical Table

Lot or serial number

0403809001 to 0425514019

Model or catalog number

ST01-410-X

Companies
Manufacturer
Steris Corporation
2720 Gunter Park East
Montgomery
36109
Alabama
UNITED STATES

B) Steris 5085SRT Surgical Table

Lot or serial number

0403809001 to 0425514019

Model or catalog number

STO1-420-X

Companies
Manufacturer
Steris Corporation
2720 Gunter Park East
Montgomery
36109
Alabama
UNITED STATES