Various Dimension Chemistry, Clinical Chemistry, EXL and XP System Products - II (2014-10-10)
- Starting date:
- October 10, 2014
- Posting date:
- November 24, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-42131
Reason
Multiple customer complaints of leaking dimension flexes while inside the flex wrapper. There has been a recent increase of complaints starting in April 2014 citing lidstock cuts and loose lidstock (delamination) resulting in the leaking flexes. There may be a single leaking flex or multiple leaking flexes from a carton as well as leaks in multiple cartons.
Recalled Products
- Dimension Clinical Chemistry System-ALP/Alkaline Phosphatase
- Dimension Clinical Chemistry System -Revised Amylase
- Dimension Chemistry System - Troponin I
- Dimension Clinical Chemistry System -ALB/Albumin
- Dimension Clinical Chemistry System -Total Iron Binding Capacity (TIBC)
- Dimension Clinical Chemistry System -Complement C3 Flex Reagent
- Dimension Clinical Chemistry System -Transferrin Reagent Cartridge
- imension Clinical Chemistry System-Enzymatic Carbonate Flex Reagent
- Dimension Clinical Chemistry System -lithium flex reagent cartridge
- Dimension Xpand System-Enzymatic Carbonate Flex Reagent
- Dimension Xpand System-Human Chorionic Gonadotropin Flex (IHCG)
- Dimension Xpand System-ITNI Cardiac Troponin I Flex Reagent
- Dimension xpand system-Immb Mass Creatine Kinase MB Isoenzyme Flex Assay
- Dimension Chemistry System -(FPSA) Free Prostate Ppecific Antigen Flex Reagent
- Dimension Chemistry System -(TPSA) Total Prostate Specific Antigen Flex Reagent
- Dimension Clinical Chemistry System -LDL Choloesterol Flex Reagent Cart. (ALDL)
- Dimension Clinical Chemistry System -Glucose Flex Reagent Cartridge (Gluc)
- Dimension Clinical Chemistry System -Urine Amphet./Methamphet. Screen Flex
- Dimension Clinical Chemistry System-Microalbumin Flex Reagent Cartridge (MALB)
- Dimension EXL System-Cardiac Troponin-I(CTNI) Assay
- Dimension EXL System -Free PSA (FPSA) Assay
- Dimension EXL System -Mass Creatine Kinase MB Isoenzyme (MMB) Assay
- Dimension EXL System -Myoglobin (MYO) Assay
- Dimension EXL System - N-terminal Pro-Brain Natriuretic Peptide (PBNP) assay
- Dimension EXL System -Tacrolimus (TACR) Assay
- Dimension EXL System -Total PSA (TPSA) assay
- Dimension EXL System - Creatine Kinase Flex (CKI) assay
- Dimension EXL System - Lactate Dehydrogenate (LDI) Assay
- Dimension XP and System- Creatine Kinase Flex (CKI) assay
- Dimension XP and System - Lactate Dehydrogenate Flex (LDI)
- Dimension Clinical Chemistry System-Creatine Kinase Flex (CKI) assay
- Dimension Clinical Chemistry System -Lactate Dehydrogenate Flex (LDI)
- Dimension Clinical Chemistry System.- automated HDL cholesterol (AHDL) assay
- Dimension XP and System-automated HDL cholesterol (AHDL) assay
- Dimension EXL System-automated HDL cholesterol (AHDL) Assay
- Dimension Chemistry Ssystem-Creatine Kinase MB (MBI) Assay
- Dimension XPL and System-Creatine Kinase MB (MBI) Assay
- Dimension EX Lsyste,- Creatine Kinase MB (MBI) Assay
- Dimension XP and System-Total Iron Binding Capacity (IBCT) Asay
- Dimension EXL System- Total Iron Binding Capacity (IBCT) Asay
- Dimension XP and System-Acetominophen (ACTM) assay
- Dimension XP and System-Albumin (ALB) Assay
- Dimension XP and system-Ethyl Alcohol (ACL) Assay
- Dimension XP and System -Automated LDL (ALDL) Cholesterol assay
- Dimension XP and System -Alanine Aminotransferase (ALT) Assay
- Dimension XP and System -Ammonia (AMON) Assay
- Dimension XP and Aystem -Urine Amphetamine/Methamphetamine (AMPH)
- Dimension XP and System -Amylase (AMY) Assay
- Dimension XP and System -Aspartate Aminotranspherase (AST) Assay
- Dimension XP and System -Urine Bariturate (BARB) Screen
- Dimension XP and System -Urine Benzodiazepines (BENZ) Screen
- Dimension XP and System -Urea Nitrogen (BUN) Assay
- Dimension XP and System -Complement C3 Assay
- Dimension XP and System -Complement C4 Assay
- Dimension XP and System-Calcium (Ca) Assay
- Dimension XP and System -Cardio Phase High Sensitivity CRP (CCRP)
- Dimension XP and System -Cholesterol (CHOL) Assay
- Dimension XP and System -Urine Cocaine (COC) Metabolite Screen
- Dimension XP and System -Carbamazapine (CRBM) Assay
- Dimension XP and System -Creatinine (CREA) Assay
Affected products
1) Dimension Clinical Chemistry System-ALP/Alkaline Phosphatase
Lot or serial number
Not applicable
Model or catalog number
- DF15A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
2) Dimension Clinical Chemistry System -Revised Amylase
Lot or serial number
Not applicable
Model or catalog number
- DF17A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
3) Dimension Chemistry System - Troponin I
Lot or serial number
Not applicable
Model or catalog number
- DF73
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
4) Dimension Clinical Chemistry System -ALB/Albumin
Lot or serial number
Not applicable
Model or catalog number
- DF13
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
5) Dimension Clinical Chemistry System -Total Iron Binding Capacity (TIBC)
Lot or serial number
Not applicable
Model or catalog number
- 10444944 and
- DF84
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
6) Dimension Clinical Chemistry System -Complement C3 Flex Reagent
Lot or serial number
Not applicable
Model or catalog number
- DF101 and
- DF102
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
7) Dimension Clinical Chemistry System -Transferrin Reagent Cartridge
Lot or serial number
Not applicable
Model or catalog number
- DF103
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
8) imension Clinical Chemistry System-Enzymatic Carbonate Flex Reagent
Lot or serial number
Not applicable
Model or catalog number
- 10444975 and
- DF137
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
9) Dimension Clinical Chemistry System -lithium flex reagent cartridge
Lot or serial number
Not applicable
Model or catalog number
- DF132
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
10) Dimension Xpand System-Enzymatic Carbonate Flex Reagent
Lot or serial number
Not applicable
Model or catalog number
- 10444972 and
- DF137
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
11) Dimension Xpand System-Human Chorionic Gonadotropin Flex (IHCG)
Lot or serial number
Not applicable
Model or catalog number
- RF530
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
12) Dimension Xpand System-ITNI Cardiac Troponin I Flex Reagent
Lot or serial number
Not applicable
Model or catalog number
- 10444905,
- RF421C and
- RF521
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
13) Dimension xpand system-Immb Mass Creatine Kinase MB Isoenzyme Flex Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444904 and
- RF520
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
14) Dimension Chemistry System -(FPSA) Free Prostate Ppecific Antigen Flex Reagent
Lot or serial number
Not applicable
Model or catalog number
- RF452
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
15) Dimension Chemistry System -(TPSA) Total Prostate Specific Antigen Flex Reagent
Lot or serial number
Not applicable
Model or catalog number
- RF451
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
16) Dimension Clinical Chemistry System -LDL Choloesterol Flex Reagent Cart. (ALDL)
Lot or serial number
Not applicable
Model or catalog number
- DF131
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
17) Dimension Clinical Chemistry System -Glucose Flex Reagent Cartridge (Gluc)
Lot or serial number
Not applicable
Model or catalog number
- DF40
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
18) Dimension Clinical Chemistry System -Urine Amphet./Methamphet. Screen Flex
Lot or serial number
Not applicable
Model or catalog number
- DF91B
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
19) Dimension Clinical Chemistry System-Microalbumin Flex Reagent Cartridge (MALB)
Lot or serial number
Not applicable
Model or catalog number
- 10444896,
- 10444905,
- RF421C and
- RF521
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
20) Dimension EXL System-Cardiac Troponin-I(CTNI) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444896,
- 10444905,
- RF421C and
- RF521
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
21) Dimension EXL System -Free PSA (FPSA) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444888 and
- RF452
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
22) Dimension EXL System -Mass Creatine Kinase MB Isoenzyme (MMB) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444895,
- 10444904,
- RF420 and
- RF520
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
23) Dimension EXL System -Myoglobin (MYO) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444907
- RF422A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
24) Dimension EXL System - N-terminal Pro-Brain Natriuretic Peptide (PBNP) assay
Lot or serial number
Not applicable
Model or catalog number
- 10444901
- RF423A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
25) Dimension EXL System -Tacrolimus (TACR) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444938
- DF107
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
26) Dimension EXL System -Total PSA (TPSA) assay
Lot or serial number
Not applicable
Model or catalog number
- 10444889
- RF451
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
27) Dimension EXL System - Creatine Kinase Flex (CKI) assay
Lot or serial number
Not applicable
Model or catalog number
- 10464511
- DF38
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
28) Dimension EXL System - Lactate Dehydrogenate (LDI) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10284483
- DF54
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
29) Dimension XP and System- Creatine Kinase Flex (CKI) assay
Lot or serial number
Not applicable
Model or catalog number
- 10464511
- DF38
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
30) Dimension XP and System - Lactate Dehydrogenate Flex (LDI)
Lot or serial number
Not applicable
Model or catalog number
- 10284483
- DF54
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
31) Dimension Clinical Chemistry System-Creatine Kinase Flex (CKI) assay
Lot or serial number
Not applicable
Model or catalog number
- 10464511 and
- DF38
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
32) Dimension Clinical Chemistry System -Lactate Dehydrogenate Flex (LDI)
Lot or serial number
Not applicable
Model or catalog number
- 10284483
- DF54
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
33) Dimension Clinical Chemistry System.- automated HDL cholesterol (AHDL) assay
Lot or serial number
Not applicable
Model or catalog number
- 10464332
- DF48B
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
34) Dimension XP and System-automated HDL cholesterol (AHDL) assay
Lot or serial number
Not applicable
Model or catalog number
- 10464332
- DF48B
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
35) Dimension EXL System-automated HDL cholesterol (AHDL) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10464332
- DF48B
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
36) Dimension Chemistry Ssystem-Creatine Kinase MB (MBI) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10464510
- DF32
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
37) Dimension XPL and System-Creatine Kinase MB (MBI) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10464510
- DF32
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
38) Dimension EX Lsyste,- Creatine Kinase MB (MBI) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10464510
- DF32
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
39) Dimension XP and System-Total Iron Binding Capacity (IBCT) Asay
Lot or serial number
Not applicable
Model or catalog number
- 10444944
- DF84
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
40) Dimension EXL System- Total Iron Binding Capacity (IBCT) Asay
Lot or serial number
Not applicable
Model or catalog number
- 10444944
- DF84
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
41) Dimension XP and System-Acetominophen (ACTM) assay
Lot or serial number
Not applicable
Model or catalog number
- 10444916
- DF88
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
42) Dimension XP and System-Albumin (ALB) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444975
- DF13
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
43) Dimension XP and system-Ethyl Alcohol (ACL) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444917
- DF18
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
44) Dimension XP and System -Automated LDL (ALDL) Cholesterol assay
Lot or serial number
Not applicable
Model or catalog number
- 10444890
- DF131
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
45) Dimension XP and System -Alanine Aminotransferase (ALT) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444958
- 10475530
- DF143
- DF43A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
46) Dimension XP and System -Ammonia (AMON) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444954
- DF19
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
47) Dimension XP and System -Urine Amphetamine/Methamphetamine (AMPH)
Lot or serial number
Not applicable
Model or catalog number
- 10444926
- DF91B
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
48) Dimension XP and System -Amylase (AMY) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444965
- DF17A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
49) Dimension XP and System -Aspartate Aminotranspherase (AST) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444959
- DF41A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
50) Dimension XP and System -Urine Bariturate (BARB) Screen
Lot or serial number
Not applicable
Model or catalog number
- 10444918
- DF96A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
51) Dimension XP and System -Urine Benzodiazepines (BENZ) Screen
Lot or serial number
Not applicable
Model or catalog number
- 10444919
- DF97A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
52) Dimension XP and System -Urea Nitrogen (BUN) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444969
- DF21
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
53) Dimension XP and System -Complement C3 Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444981
- DF101
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
54) Dimension XP and System -Complement C4 Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444981
- DF102
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
55) Dimension XP and System-Calcium (Ca) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444949
- DF23A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
56) Dimension XP and System -Cardio Phase High Sensitivity CRP (CCRP)
Lot or serial number
Not applicable
Model or catalog number
- 10444897
- RF434
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
57) Dimension XP and System -Cholesterol (CHOL) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444891
- DF27
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
58) Dimension XP and System -Urine Cocaine (COC) Metabolite Screen
Lot or serial number
Not applicable
Model or catalog number
- 10444920
- DF92A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
59) Dimension XP and System -Carbamazapine (CRBM) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444930
- DF87
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
60) Dimension XP and System -Creatinine (CREA) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444968
- DF33A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES