Various Dimension Chemistry, Clinical Chemistry, EXL and XP System Products - IV (2014-10-10)
- Starting date:
- October 10, 2014
- Posting date:
- November 24, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-42145
Reason
Multiple customer complaints of leaking dimension flexes while inside the flex wrapper. There has been a recent increase of complaints starting in April 2014 citing lidstock cuts and loose lidstock (delamination) resulting in the leaking flexes. There may be a single leaking flex or multiple leaking flexes from a carton as well as leaks in multiple cartons.
Recalled Products
- Dimension EXL System - C-Reactive Protein (CRP) Assay
- Dimension EXL System - Direct Bilirubin (DBI) Assay
- Dimension EXL System - Digoxin (DGNA) Assay
- Dimension EXL System - Ethyl Alcohol (EtOH) Assay
- Dimension EXL System - Urine Ecstacy (EXTC) Screen
- Dimension EXL System - Ferritin (FERR) Assay
- Dimension EXL System - Gentamicin (GENT) Assay
- Dimension EXL System - Gamma-Glutamyl Transferase (GGT) Assay
- Dimension EXL System - Glucose (GLUC) Assay
- Dimension EXL System - Human Chorionic Gonadotropin (HCG) Assay
- Dimension EXL System - Immunoglobulin A (IgA) assay
- Dimension EXL System - Immunoglobulin G(IgG) Assay
- Dimension EXL System - Immunoglobulin M (IgM) Assay
- Dimension EXL System - Iron (Iron) Assay
- Dimension EXL System - Lactic Acid (LA) Assay
- Dimension EXL System - Lithium (LI) assay
- Dimension EXL System - Lipase (LIPL) Assay
- Dimension EXL System - Microalbumin (MALB) Assay
- Dimension EXL System - Urine Methadone (METH) Screen
- Dimension EXL System - Urine Opiates (OPI) Screen
- Dimension EXL System - Prealbumin (PALB) Assay
- Dimension EXL System - Pseudocholinesterase (PCHE) Assay
- Dimension EXL System - Urine Phencyclidine (PCP) Screen
- Dimension EXL System - Phenobarbital (PHNO) Assay
- Dimension EXL System - Phosphorus (PHOS) Assay
- Dimension EXL System - Phenytoin (PTN) Assay
- Dimension EXL System - c-reactive protein extended range (rcrp) assay
- Dimension EXL System - Salicylate (SAL) Assay
- Dimension EXL System - Thyroxine (T4) Assay
- Dimension EXL System - Total Bilirubin (TBIL) Assay
- Dimension EXL System - Triglycerides (TGL) Assay
- Dimension EXL System - Urine Cannabinoids (THC) Screen
- Dimension EXL System - Theophyline (THEO) Assay
- Dimension EXL System - Tobramycin (TOBR) Assay
- Dimension EXL System - Total Protein (TP) Assay
- Dimension EXL System - Transferrin (TRNF) Assay
- Dimension EXL System - Urine/Cerebrospinal Fluid Protein (UCFP) Assay
- Dimension EXL System - Uric Acid (URCA) Assay
- Dimension EXL System - Valproic Acid (VALP) Assay
- Dimension EXL System - Vancomycin (VANC) Assay
- Dimension Chemistry System - Sirolimus (Siro) Assay
- Dimension XP and System - Sirolimus (SIRO) Assay
- Dimension EXL System - Sirolimus (SIRO) Assay
- Dimension XP and System - Myoglobin (MYO) Assay
- Dimension XP and System - Total Prostate Specific Antigen (TPSA) Assay
- Dimension XP and System - Free Prostate Specific Antigen (FPSA) Assay
- Dimension XP and System - Alkaline Phosphatase (ALP) Assay
- Dimension XP and System - Magnesium (Mg) Assay
- Dimension EXL System - Alkaline Phosphatase (ALP) Assay
- Dimension EXL System - Magnesium (Mg) Assay
- Dimension EXL System - Hemoglobin A1C (Hb1C) Assay
- Dimension XP and System - Hemoglobin A1C (Hb1C) Assay
- Dimension Clinical Chemistry System - Hemoglobin A1C (Hb1C) Assay
- Dimension Clinical Chemistry System - Alkaline Phosphatase (ALPI) Assay
- Dimension EXL System - Alkaline Phosphatase (ALPI) Assay
- Dimension EXL System - Enzymatic Carbonate Flex Reagent Cartridge (ECO2)
- Dimension EXL System - Ammonia (AMM) Assay
- Dimension XP and System - Ammonia (AMM) Assay
- Dimension Clinical Chemistry System - Ammonia (AMM) Assay
- Dimension XP and System - Alkaline Phosphatase (ALPI) Assay
Affected products
1) Dimension EXL System - C-Reactive Protein (CRP) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444894
- DF37
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
2) Dimension EXL System -Direct Bilirubin (DBI) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444956
- DF125
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
3) Dimension EXL System -Digoxin (DGNA) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444941
- DF35A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
4) Dimension EXL System -Ethyl Alcohol (EtOH) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10459408
- DF22
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
5) Dimension EXL System -Urine Ecstacy (EXTC) Screen
Lot or serial number
Not applicable
Model or catalog number
- 10444921
- DF109
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
6) Dimension EXL System - Ferritin (FERR) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444946
- RF440
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
7) Dimension EXL System - Gentamicin (GENT) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444927
- DF12
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
8) Dimension EXL System - Gamma-Glutamyl Transferase (GGT) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444960
- DF45A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
9) Dimension EXL System - Glucose (GLUC) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444971
- DF40
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
10) Dimension EXL System - Human Chorionic Gonadotropin (HCG) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444913
- 10444914
- RF430
- RF530
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
11) Dimension EXL System-Immunoglobulin A (IgA) assay
Lot or serial number
Not applicable
Model or catalog number
- 10444982
- DF74
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
12) Dimension EXL System - Immunoglobulin G(IgG) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444983
- DF76
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
13) Dimension EXL System - Immunoglobulin M (IgM) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444984
- DF81
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
14) Dimension EXL System - Iron (Iron) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444945
- DF85
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
15) Dimension EXL System -Lactic Acid (LA) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444961
- DF16
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
16) Dimension EXL System - Lithium (LI) assay
Lot or serial number
Not applicable
Model or catalog number
- 10444939
- DF132
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
17) Dimension EXL System - Lipase (LIPL) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10460277
- DF56
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
18) Dimension EXL System - Microalbumin (MALB) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444977
- DF114
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
19) Dimension EXL System -Urine Methadone (METH) Screen
Lot or serial number
Not applicable
Model or catalog number
- 10444922
- DF90A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
20) Dimension EXL System -Urine Opiates (OPI) Screen
Lot or serial number
Not applicable
Model or catalog number
- 10444923
- DF93A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
21) Dimension EXL System - Prealbumin (PALB) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444978
- DF80
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
22) Dimension EXL System - Pseudocholinesterase (PCHE) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444964
- DF51
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
23) Dimension EXL System -Urine Phencyclidine (PCP) Screen
Lot or serial number
Not applicable
Model or catalog number
- 10444924
- DF94A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
24) Dimension EXL System - Phenobarbital (PHNO) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444933
- DF60
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
25) Dimension EXL System - Phosphorus (PHOS) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444951
- DF61
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
26) Dimension EXL System -Phenytoin (PTN) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444940
- DF64
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
27) Dimension EXL System - c-reactive protein extended range (rcrp) assay
Lot or serial number
Not applicable
Model or catalog number
- 10444903
- DF34
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
28) Dimension EXL System -Salicylate (SAL) Assay
Lot or serial number
Not applicable
Model or catalog number
- DF20
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
29) Dimension EXL System - Thyroxine (T4) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444908
- DF65
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
30) Dimension EXL System -Total Bilirubin (TBIL) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444966
- DF67A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
31) Dimension EXL System - Triglycerides (TGL) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444906
- DF69A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
32) Dimension EXL System - Urine Cannabinoids (THC) Screen
Lot or serial number
Not applicable
Model or catalog number
- 10444925
- DF95A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
33) Dimension EXL System - Theophyline (THEO) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444942
- DF71
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
34) Dimension EXL System-Tobramycin (TOBR) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444928
- DF14
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
35) Dimension EXL System - Total Protein (TP) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444979
- 10444889
- DF73
- RF451
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
36) Dimension EXL System - Transferrin (TRNF) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444985
- DF103
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
37) Dimension EXL System - Urine/Cerebrospinal Fluid Protein (UCFP) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444950
- DF26
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
38) Dimension EXL System - Uric Acid (URCA) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444967
- DF77
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
39) Dimension EXL System - Valproic Acid (VALP) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444943
- DF78
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
40) Dimension EXL System - Vancomycin (VANC) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444929
- DF86
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
41) Dimension Chemistry System - Sirolimus (Siro) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10464331
- DF306
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
42) Dimension XP and System - Sirolimus (SIRO) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10464331
- DF306
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
43) Dimension EXL System - Sirolimus (SIRO) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10464331
- DF306
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
44) Dimension XP and System - Myoglobin (MYO) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444907
- RF422A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
45) Dimension XP and System - Total Prostate Specific Antigen (TPSA) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444889
- RF451
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
46) Dimension XP and System - Free Prostate Specific Antigen (FPSA) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444888
- RF452
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
47) Dimension XP and System - Alkaline Phosphatase (ALP) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444953
- DF15A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
48) Dimension XP and System - Magnesium (Mg) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444963
- DF57
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
49) Dimension EXL System - Alkaline Phosphatase (ALP) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444953
- DF15A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
50) Dimension EXL System - Magnesium (Mg) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10444963
- DF57
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
51) Dimension EXL System - Hemoglobin A1C (Hb1C) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10483822
- DF105A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
52) Dimension XP and System - Hemoglobin A1C (Hb1C) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10483822
- DF105A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
53) Dimension Clinical Chemistry System - Hemoglobin A1C (Hb1C) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10483822
- DF105A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
54) Dimension Clinical Chemistry System - Alkaline Phosphatase (ALPI) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10642445
- DF150
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
55) Dimension EXL System - Alkaline Phosphatase (ALPI) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10642445
- DF150
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
56) Dimension EXL System - Enzymatic Carbonate Flex Reagent Cartridge (ECO2)
Lot or serial number
Not applicable
Model or catalog number
- 10444972
- DF137
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
57) Dimension EXL System - Ammonia (AMM) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10711991
- DF119
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
58) Dimension XP and System - Ammonia (AMM) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10711991
- DF119
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
59) Dimension Clinical Chemistry System - Ammonia (AMM) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10711991
- DF119
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
60) Dimension XP and System - Alkaline Phosphatase (ALPI) Assay
Lot or serial number
Not applicable
Model or catalog number
- 10642445
- DF150
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES