This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
REFLECTION 3 Acetabular System-Constrained Acetabular Liner (2015-05-08)
- Starting date:
- May 8, 2015
- Posting date:
- June 3, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-53657
Recalled Products
A. REFLECTION 3 Acetabular System-Constrained Acetabular Liner
Reason
Smith & Nephew has initiated a product recall of select lots of the R3 constrained acetabular liners due to a manufacturing error. The R3 constrained outer liner locking groove diameter is 0.030" oversized and may not allow the locking ring to engage into the R3 shell.
Affected products
A. REFLECTION 3 Acetabular System-Constrained Acetabular Liner
Lot or serial number
- 14HM08141
- 14HM08148
- 14HM08176
Model or catalog number
- 71339160
- 71339162
Companies
- Manufacturer
-
Smith & Nephew Inc.
1450 Brooks Road
Memphis, Tennessee
38116
UNITED STATES