Dialog+ SW 9.XX (2015-11-18)
- Starting date:
- November 18, 2015
- Posting date:
- January 7, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-56560
Affected Products
A. Dialog+ SW 9.XX
Reason
Potential problem if servicing device and interchanging tubing between valve block level regulation and diaphragm pump if the functionality of the level regulation system is not tested as requested in the service manual. An interchanged tubing results in an inverse function of the manual level regulation system.
Affected products
A. Dialog+ SW 9.XX
Lot or serial number
All lots.
Model or catalog number
710200CV
Companies
- Manufacturer
-
B. Braun Avitum Ag
Schwarzenberger Weg 73-79
Melsungen
34212
GERMANY