This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
ENTERALITE INFINITY ENTERAL FEEDING SETS (2015-12-01)
- Starting date:
- December 1, 2015
- Posting date:
- January 14, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-56600
Affected Products
A. ENTERALITE INFINITY BAG SET
B. Enteralite Infinity Spike Set
Reason
The manufacturer has become aware of an increase in the number of complaints for EnterLite Infinity Enteral Pump Delivery sets with ENFit connector system and transitional stepped connectors (administration sets) that are used with the EnteraLite Infinity Enteral Feeding Pump. Customers are reporting formula leaking between the purple ENFit connector and the white transitional stepped connector due to cracking or breaking of the purple connector.
Affected products
A. ENTERALITE INFINITY BAG SET
Lot or serial number
All lots.
Model or catalog number
- INF0500-A
- INF0500-A
- INF1200-A
- INF1200-A
Companies
- Manufacturer
-
Zevex Inc.
also trading as Moog Medical Devices Group
4314 Zevex Park Lane
Salt Lake City
84123
Utah
UNITED STATES
B. Enteralite Infinity Spike Set
Lot or serial number
All lots.
Model or catalog number
- INF0020-A
- INF0020-A
Companies
- Manufacturer
-
Zevex Inc.
also trading as Moog Medical Devices Group
4314 Zevex Park Lane
Salt Lake City
84123
Utah
UNITED STATES