Health product recall

PNEUMO SURE HIGH FLOW INSUFFLATOR - MAIN UNIT (2016-03-28)

Starting date:
March 28, 2016
Posting date:
April 22, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58028

Affected products

A. PNEUMO SURE HIGH FLOW INSUFFLATOR - MAIN UNIT

Reason

When operating the device in high flow mode (40L/min, 15 mmHg) and connecting to a house gas inlet source, within seconds of insufflating the flow rate would drop to 0L/min and the red "check gas supply" symbol would appear on the display, immediately suspending insufflation.

Affected products

A. PNEUMO SURE HIGH FLOW INSUFFLATOR - MAIN UNIT

Lot or serial number

1101CE178

Model or catalog number

620-040-610

Companies
Manufacturer
Stryker Endoscopy
5900 Optical Court
San Jose
California
UNITED STATES