This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

EMPTY,STERILE I.V.,SOLUTION TRANSFER CONTAINERS (2016-06-29)

Starting date:
June 29, 2016
Posting date:
July 19, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-59418

Affected products

A. EMPTY,STERILE I.V.,SOLUTION TRANSFER CONTAINERS

Reason

AutoComp6, AutoComp6- XP and AutoComp6-XPS high speed compounders may not be able to maintain the fill accuracy requirements for the compounded bags as outlined in the user's manual. Therefore, the patient receiving the compounded bag may both receive the correct quantity of individual parenteral nutrition solutions or the correct overall quantity of parenteral nutrition solutions.

Affected products

A. EMPTY,STERILE I.V.,SOLUTION TRANSFER CONTAINERS

Lot or serial number

All lots

Model or catalog number
  • "58740
  • 58741
  • 58742
  • 58750
  • 58753
  • 58754
  • 58781
  • 58785
  • 66005
  • 66010
  • 66015
  • 66020
  • 66025
  • 66030
  • 66035
  • 66310
  • 66315
  • 66325
  • 66330
  • 66710
  • 66715"
Companies
Manufacturer
THE METRIX COMPANY
4400 CHAVENELLE RD
DUBUQUE
52002
Iowa
UNITED STATES