Health product recall

Volista Surgical Lights (2018-01-08)

Starting date:
January 8, 2018
Posting date:
January 29, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65780

Affected products

Volista Surgical Lights

Reason

There is a possibility that the central handle holder detaches from the cupola during use. We determined that the root cause of this malfunction is excessive tightening of the screws during mounting of the handle holder, resulting in the rivets loosening and detaching from the central handle holder ring. There is a remote chance that this issue may cause post-operative infection, foreign body granuloma or allergic reactions, if the particles fall into the patient's wound or on sterile equipment.

Affected products

Volista Surgical Lights

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • ARD568811903
  • ARD568811912
  • ARD568811913
  • ARD568811952
  • ARD568811961
  • ARD568811963
  • ARD568812912
  • ARD568812913
  • ARD568812952
  • ARD568812963
  • ARD568821960
  • ARD568821961
Companies
Manufacturer

Maquet S.A.

Avenue de la Pomme De Pin, Parc De Limere

Ardon, Orleans

45074

FRANCE