Health product recall

Somatom Emotion 16-Slice Configuration and Somatom Perspective - CT Scanner (2018-02-05)

Starting date:
February 5, 2018
Posting date:
March 13, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66124

Affected products

  1. SOMATOM EMOTION 16-SLICE CONFIGURATION
  2. SOMATOM PERSPECTIVE - CT SCANNER

Reason

Siemens healthcare became aware of possible incorrect tube current calculations by the Care Dose4d algorithm for head scans based on p.a. (posterior-anterior) or a.p. topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure.