Health product recall

Permobil Powered Wheelchairs (2018-08-06)

Starting date:
August 6, 2018
Posting date:
August 17, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67446

Affected products

Permobil Powered Wheelchairs

Reason

Backrest support telescoping bars potentially fail in certain instances. The backrest support telescoping bars allow for the adjustment of the backrest height. Reports received consist of 1 or both back canes on the PS Backrest Frame Assembly, having broken while end-user is seated in the device. When this reported failure occurs, reports indicate the partial or complete detachment of the back frame from the seating which allows the end-user to lose positioning and fall backwards.

Affected products

Permobil Powered Wheelchairs

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • PS C300
  • PS C350
  • PS C400
  • PS C500
  • PS M300
Companies
Manufacturer
PERMOBIL INC
300 DUKE DRIVE
LEBANON
37090
Tennessee
UNITED STATES