Health product recall

Alinity i Free T3 Reagent Kit and Alinity I Total T3 Reagent Kit (2018-10-16)

Starting date:
October 16, 2018
Posting date:
November 2, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68174

Affected products

  1. Alinity i Free T3 Reagent Kit
  2. Alinity i Total T3 Reagent Kit

Reason

During Non-Conformance investigation a Reagent Cross Contamination (RCC) has been observed for the 6 and 2-

point ARCHITECT Free T3 and Total T3 assays on the i1000, when tested alongside assays containing poly-l-lysine (pll) e.g. ARCHITECT TSH. RCC has also been observed for the 2 and 6-point ARCHITECT free T3 and total T3 assays when tested on the i2000/i2000SR system and the Alinity i System.

Affected products

A. Alinity i Free T3 Reagent Kit

Lot or serial number

All lots.

Model or catalog number
  • 07P6920
  • 07P6930
Companies
Manufacturer

Abbott Ireland Diagnostic Division

Lisnamuck, Longford

IRELAND


B. Alinity i Total T3 Reagent Kit

Lot or serial number

All lots.

Model or catalog number
  • 07P9420
  • 07P9430
Companies
Manufacturer

Abbott Ireland Diagnostic Division

Lisnamuck, Longford

IRELAND