CARESCAPE Central Station (2018-01-10)
- Starting date:
- January 10, 2018
- Posting date:
- January 9, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68636
Affected prodcuts
CARESCAPE Central Station
Reason
GE Healthcare has recently become aware that after boot-up or system start up, CSCS v2.0.2 units may display unexpected no comm notifications in the patient multi-viewer or single viewer window when the CSCS is connected to both the mission critical (mc) and the information exchange (ix) networks. If a no comm notification is displayed, all patient waveforms, parameter data and alarms are lost at the CSCS and patients admitted to telemetry transmitters are unmonitored unless monitored via other central stations, bedside monitors or direct clinical observation.
Affected products
CARESCAPE Central Station
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
2051117-005
Companies
- Manufacturer
-
GE Medical Systems Information Technologies, Inc.
8200 W Tower Ave
Milwaukee
53223
Wisconsin
UNITED STATES