DUREX Real Feel Extra Lubricated 10ct condoms (2018-07-30)

Starting date:
July 30, 2018
Posting date:
August 17, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68726

Affected products

DUREX Real Feel Extra Lubricated 10ct condoms

Reason

Following observation of variable burst pressure performance of polyisoprene condoms manufactured during 2018; Reckitt Benckise conducted an internal investigation and identified that one batch of Non-Latex Durex RealFeel Extra Lubricated 10ct Condoms, UPC: 0 67981 98715 7, Batch: 1000443254, Expiry Date: 2023 - 02 marketed in Canada between June 2018 to July 2018 is not expected to meet the internal burst pressure specification of 1.0 kPa at the end of 5 year shelf-life. Consequently, Reckitt Benckiser (Canada) Inc. as a cautious approach is initiating a consumer level recall.

Affected products

DUREX Real Feel Extra Lubricated 10ct condoms

Lot or serial number

1000443254

Model or catalog number

98715