Nokia BPM+ (2018-12-17)

Starting date:
December 17, 2018
Posting date:
January 23, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68886

Affected products

Nokia BPM+

Reason

BPM+ is an upper arm blood pressure monitor that measures systolic and diastolic pressure, and heart rate. It is intended to be used at home by an adult with an arm circumference ranging from 22 to 42 cm. It was identified that the device was less accurate on subjects with larger arm circumference (greater then 35 cm).

Affected products

Nokia BPM+

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

BP-801

Companies
Manufacturer
Withings
2 Rue Maurice Hartmann
Issy-les-Moulineaux
92130
FRANCE