Health product recall

Achieva 1.5T, 3.0T, XR MRI Systems, Intera 1.5T, 3.0T MRI Systems, Ingenia 1.5T, 3.0T, 3.0T CX, Ingenia Elition X, and Panorama HFO (2018-12-17)

Starting date:
December 17, 2018
Posting date:
March 15, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69320

Affected Products

  1. Achieva 3.0T MRI System - Main Unit
  2. Achieva 1.5T MRI System - Main Unit
  3. Achieva XR System - Main Unit
  4. Panorama HFO
  5. Intera 1.5T MRI System - Main Unit
  6. Intera 3.0T MRI System
  7. Ingenia 1.5T
  8. Ingenia 3.0T
  9. Ingenia 3.0T CX
  10. Ingenia Elition X

Reason

Philips MR systems are standard equipped with two ceiling speakers, positioned in the front and/or rear of the MR system. These speakers enable communication between the operator in the examination room and the patient and offer stereo music or ambient experience audio. These speakers are in most cases fixated in the suspended ceiling of the MR examination room according the Philips service instructions. The Philips Planning Reference Data (PRD) prescribes the suspended ceiling must be strong enough to hold 25 N speakers.

However in very rare cases, due to external influences (hospital facility maintenance, water leakage, degradation of speaker fixation) during the lifetime of the MR system, the ceiling speaker(s) could fall from their position and be pulled to the MR system by the magnetic field.

Affected products

A. Achieva 3.0T MRI System - Main Unit

Lot or serial number

Not applicable.

Model or catalog number

ACHIEVA 3.0T

Companies
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS

B. Achieva 1.5T MRI System - Main Unit

Lot or serial number

Not applicable.

Model or catalog number

ACHIEVA 1.5T

Companies
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS

C. Achieva XR System - Main Unit

Lot or serial number

Not applicable.

Model or catalog number

ACHIEVA XR

Companies
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS

D. Panorama HFO

Lot or serial number

Not applicable.

Model or catalog number

PANORAMA HFO

Companies
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS

E. Intera 1.5T MRI System - Main Unit

Lot or serial number

Not applicable.

Model or catalog number

INTERA 1.5T

Companies
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS

F. Intera 3.0T MRI System

Lot or serial number

Not applicable.

Model or catalog number

INTERA 3.0T

Companies
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS

G. Ingenia 1.5T

Lot or serial number

Not applicable.

Model or catalog number

1.5T

Companies
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS

H. Ingenia 3.0T

Lot or serial number

Not applicable.

Model or catalog number

3.0T

Companies
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS

I. Ingenia 3.0T CX

Lot or serial number

Not applicable.

Model or catalog number

781271

Companies
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS

J. Ingenia Elition X

Lot or serial number

Not applicable.

Model or catalog number

781358

Companies
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS