Oxygo Portable Oxygen Concentrator (2019-08-16)

Starting date:
August 16, 2019
Posting date:
August 30, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70853

Last updated:

Affected Products

Oxygo Portable Oxygen Concentrator

Reason

A potential abrupt shut down of the oxygen concentrator may occur due to an over-draw when a fully charged 8-cell battery is connected to the oxygen concentrator as its only power source. The oxygen concentrator may become unresponsive for up to one minute, after which it will behave as though a battery was just connected with standard flash screen on display.

Affected products

Oxygo Portable Oxygen Concentrator

Lot or serial number

SN's: 19223909-19256614

Model or catalog number

1400-1000

Companies
Manufacturer
Applied Home Healthcare Equipment
28825 Ranney Parkway
Westlake
44145
Ohio
UNITED STATES