Health product recall

Genius 2 & 3 Tympanic Thermometer (2019-10-25)

Starting date:
October 25, 2019
Posting date:
November 22, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71666

Last updated: 2019-11-22

Affected Products

  1. Genius 2 Tympanic Thermometer
  2. Genius 3 Tympanic Thermometer

Reason

The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (0.2°C for Genius 2 and 0.3°C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2°C for Genius 2 or +0.3°C for Genius 3.

Affected products

  1. Genius 2 Tympanic Thermometer

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

303000

Companies
Manufacturer
Cardinal Health 200, LLC (carrying on the patient recovery business from Covidien LLC)
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES

B. Genius 3 Tympanic Thermometer

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number

303013

Companies
Manufacturer
Cardinal Health 200, LLC (carrying on the patient recovery business from Covidien LLC)
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES