Health product recall

Stainless Steel Reprocessing Tray, 8mm Endoscope, IS4000 and 2D Camera Head, IS4000 (2019-12-19)

Starting date:
December 19, 2019
Posting date:
January 10, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72083

Last updated: 2020-01-10

Affected Products

  1. Stainless Steel Reprocessing Tray, 8mm Endoscope, IS4000
  2. Stainless Steel Reprocessing Tray, 2D Camera Head, IS4000

Reason

Intuitive has become aware of, via field complaints, the potential for small, but detectable holes to be generated in the sterilization wrap used with Da Vinci 8 mm endoscope sterilization trays (PN 400498-02) and handheld camera sterilization trays (PN 400499-03) which could lead to a breach in sterility.

Affected products

  1. Stainless Steel Reprocessing Tray, 8mm Endoscope, IS4000

Lot or serial number

All lots.

Model or catalog number

400498-02

Companies
Manufacturer
Intuitive Surgical Inc.
1266 Kifer Road, Building 100, 101, 102, 103
Sunnyvale
94086
California
UNITED STATES

B. Stainless Steel Reprocessing Tray, 2D Camera Head, IS4000

Lot or serial number

All lots. 

Model or catalog number

400499-03

Companies
Manufacturer
Intuitive Surgical Inc.
1266 Kifer Road, Building 100, 101, 102, 103
Sunnyvale
94086
California
UNITED STATES