Health product recall

Change Healthcare Cardiology ECG Management System (2020-02-09)

Starting date:
February 9, 2020
Posting date:
March 13, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72533

Last updated: 2020-03-13

Affected Products

Change Healthcare Cardiology ECG Management System

Reason

Users can view the ECG waveforms using either simultaneous or sequential ECG format options, depending on user configuration. By default, the simultaneous ECG format is presented on a scale of 25 mm/sec whereas in the sequential view the scale dynamically changes according to the resolution of the monitor in use.

Change Healthcare has identified an issue whereby users who are estimating heart rate by counting grid squares, may assume that the sequential ECG format is presented on a scale of 25 mm/sec and thereby may inaccurately estimate the heart rate value.

Inaccurate estimation of the heart rate value may occur in instances where the user isestimating heart rate counting grid squares using the sequential ECG format. The user may assume that the ECG is presented on a scale of 25 mm/sec and thereby may inaccurately estimate the heart rate value.

Affected products

Change Healthcare Cardiology ECG Management System

Lot or serial number

Version: 13.2.1 HF2

Model or catalog number

Version 14.X

Companies
Manufacturer
Change Healthcare Canada Company
10711 Cambie Rd.
Richmond
V6X 3G5
British Columbia
CANADA