Health product recall

ITS4850A Intellivue Telemetry System (2020-02-28)

Starting date:
February 28, 2020
Posting date:
March 13, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72589

Last updated: 2020-03-13

Affected Products

ITS4850A Intellivue Telemetry System

Reason

The ECG signal from patients being monitored using a Philips TRx4851A Telemetry Transceiver may not be properly analyzed when the transceiver is used with a Philips Patient Information Center iX Release C.02.00, C.02.02, C.02.03 (all released versions of C.02, collectively referred to as ‘C.02.xx’) or C.03.01. If this occurs, the Information Center will not display a heart rate or generate, display or annunciate any heart rate or arrhythmia alarms.

Alarming based on pulse oximetry (SpO2) signals are unaffected by this issue.